Ethicon, Inc.: Medical Device Recall in 2018 - (Recall #: Z-2462-2018)
See the recall detail below. You can also see other recalls from the same firm in 2018.
PROLENE SUTURE 36"(90CM) 3-0 BLU, D9844
Class II
The needle associated with certain suture product codes has the potential to perforate the packaging due to a packaging design issue. A perforated package has the potential to cause a soft tissue injury to a healthcare professional handling the subject product or a patient infection due to a breach of sterility within the packaging if it goes unnoticed by the healthcare professional.
KDQ180 KJH311 KKH758 KMP871 KPH811 LAH127 LAH648 LDP694 LHB945 LJE966 LMJ318 LMR835 MAH349 MBE216 MCQ994 MCQ537 MCQ538
Accounts in AZ, CA, CO, FL, GA, IL, IN, KS, MA, MN, MO, NC, NE, NJ, NY, OH, OR, SC, SD, TN, TX, VA, WA, WI, WV.
Voluntary: Firm initiated