Ethicon, Inc.: Medical Device Recall in 2022 - (Recall #: Z-0480-2023)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2022.
Data Source: FDA.
Product Description:
PDS II (polydioxanone) Suture- VIOLET MONOFILAMENT 27IN 0 S/A CT-1 Product Code: Z340H For use in soft tissue approximation, including use in pediatric cardiovascular tissue where growth is expected to occur and ophthalmic surgery
Product Classification:
Class II
Date Initiated: November 16, 2022
Date Posted: December 21, 2022
Recall Number: Z-0480-2023
Event ID: 91192
Reason for Recall:
knot tensile strengths were observed to be significantly below the minimum knot tensile, will result in breakage prior to use or intra-operatively rather than premature post-operative breakage
Status: Ongoing
Product Quantity: 7,560 eaches
Code Information:
UDI-DI: 10705031060623 (individual unit); 30705031060634 (sales unit box) Lot: SCMSHK Exp. Date: February 28, 2027
Distribution Pattern:
International (OUS) only: Canada, Japan, Australia and New Zealand
Voluntary or Mandated:
Voluntary: Firm initiated