ev3 Inc.: Medical Device Recall in 2022 - (Recall #: Z-0450-2022)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2022.

Data Source: FDA.
Product Description:

Medtronic HawkOne Directional Atherectomy System, Catalog numbers H1-M and H1-M-INT., Multi-vessel, 6F (2.2 mm) Recommended sheath, 0.014" (0.36 mm) Compatible guidewire. The HawkOne directional atherectomy system is intended for use in atherectomy of the peripheral vasculature. The HawkOne catheter is indicated for use in conjunction with the SpiderFX embolic protection device in the treatment of severely calcified lesions. The HawkOne catheter is not intended for use in the coronary, carotid, iliac, or renal vasculature.

Product Classification:

Class I

Date Initiated: December 6, 2021
Date Posted: January 26, 2022
Recall Number: Z-0450-2022
Event ID: 89229
Reason for Recall:

Reports have been received of tip damage during use of the devices which resulted in some instances of tip detachment and embolization.

Status: Terminated
Product Quantity: 71,260 units
Code Information:

H1-M - Irvine GTIN 00643169792302, Galway GTIN 00643169968332. H1-M-INT - Irvine GTIN 00643169728776, Galway, GTIN 00643169968523.

Distribution Pattern:

Worldwide Distribution: US (nationwide), including PR; and Foreign distribution to countries of: Argentina, Australia, Austria, Belgium, Brazil, Chile, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Jordan, Korea, Republic Of Macao, Netherlands, New Zealand, Norway, Oman, Peru, Poland, Portugal, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Turkmenistan, United Arab Emirates, and United Kingdom.

Voluntary or Mandated:

Voluntary: Firm initiated