Event Medical LTD: Medical Device Recall in 2015 - (Recall #: Z-0230-2016)
See the recall detail below. You can also see other recalls from the same firm in 2015.
Inspiration Ventilator System, i Series and LS 12.1 LCD. The Inspiration 7i / 5i Ventilator System is to be used by healthcare professionals in hospitals or healthcare facilities and intra-hospital transport.
Class I
A component on the systems power board can fail in a manner which can cause the ventilator to shut down without alarming.
All models manufactured prior to 2015 are affected
Worldwide Distribution - US (Nationwide) and Internationally to: Algeria Argentina Australia Austria Bahrain Bangladesh Barbados Belarus Belgium Bolivia Canada Chile China Colombia Czech Rep Ecuador Egypt England Estonia Finland Germany Ghana Gibraltar Greece Guatemala Holland Hong Kong Hungary India Indonesia Iran Iraq Ireland Israel Italy Japan Jordan Kenya Korea Kuwait Laos Latin America Lebanon Libya Lithuania Malaysia Mexico Morocco Nepal Netherlands New Zealand Norway Pakistan Palestine Panama Peru Philippines Poland Portugal Russia Saudi Arabia Serbia Singapore South Africa Spain Sri Lanka Sudan Switzerland Syria Taiwan Thailand Turkey UAE Ukraine USA Venezuela Vietnam Yemen Zimbabwe
Voluntary: Firm initiated