EVIDENT SCIENTIFIC INC: Medical Device Recall in 2024 - (Recall #: Z-2416-2024)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2024.

Data Source: FDA.
Product Description:

SlideView VS200 slide viewer,

Product Classification:

Class II

Date Initiated: July 22, 2024
Date Posted: August 7, 2024
Recall Number: Z-2416-2024
Event ID: 95028
Reason for Recall:

A broken fuse inside the laser that was caused by a short circuit at PIN 4 of the DOOR CTRL. This short circuit was caused by a badly positioned cable that made contact with the housing/frame and caused a short circuit and destroyed the fuse in the laser. This caused the shutter to remain in its most recent state; either in an open or closed position. Once the fuse was destroyed the shutter could not be operated.

Status: Ongoing
Product Quantity: N/A
Code Information:

65159 67105 67104 69153 67106 69154

Distribution Pattern:

US

Voluntary or Mandated:

FDA Mandated