Exact Sciences Corporation: Medical Device Recall in 2017 - (Recall #: Z-2219-2017)
See the recall detail below. You can also see other recalls from the same firm in 2017.
Exact Sciences Sample Mixer 2. For In Vitro Diagnostic Use. The Exact Sciences Sample Mixer 2 is intended for mixing stool samples in a Sample Container as part of pre-analytic sample processing for in vitro diagnostic tests
Class II
Exact Sciences is correcting all Sample Mixer v2 Upgrade units because of incidents of Sample Mixer door damage, which has occurred when the Sample Holder opens during a run and the Sample Holder lid makes contact with the Sample Mixer door.
Equipment ID / Serial Number: MIX-0051 / MX00002051, MIX-0052 / MX00002032, MIX-0053 / MX00002028, MIX-0054 / MX00002052, MIX-0055 / MX00002043, MIX-0056 / MX00002040, MIX-0057 / MX00002050, MIX-0058 / MX00002036, MIX-0059 / MX00002027, MIX-0060 / MX00002015, MIX-0061 / MX00002016, MIX-0063 / MX00002039, MIX-0064 / MX00002033, MIX-0065 / MX00002053, MIX-0066 / MX00002014,
US to WI only
Voluntary: Firm initiated