Exact Sciences Corporation: Medical Device Recall in 2017 - (Recall #: Z-2219-2017)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2017.

Data Source: FDA.
Product Description:

Exact Sciences Sample Mixer 2. For In Vitro Diagnostic Use. The Exact Sciences Sample Mixer 2 is intended for mixing stool samples in a Sample Container as part of pre-analytic sample processing for in vitro diagnostic tests

Product Classification:

Class II

Date Initiated: April 28, 2017
Date Posted: June 7, 2017
Recall Number: Z-2219-2017
Event ID: 77260
Reason for Recall:

Exact Sciences is correcting all Sample Mixer v2 Upgrade units because of incidents of Sample Mixer door damage, which has occurred when the Sample Holder opens during a run and the Sample Holder lid makes contact with the Sample Mixer door.

Status: Terminated
Product Quantity: 15
Code Information:

Equipment ID / Serial Number: MIX-0051 / MX00002051, MIX-0052 / MX00002032, MIX-0053 / MX00002028, MIX-0054 / MX00002052, MIX-0055 / MX00002043, MIX-0056 / MX00002040, MIX-0057 / MX00002050, MIX-0058 / MX00002036, MIX-0059 / MX00002027, MIX-0060 / MX00002015, MIX-0061 / MX00002016, MIX-0063 / MX00002039, MIX-0064 / MX00002033, MIX-0065 / MX00002053, MIX-0066 / MX00002014,

Distribution Pattern:

US to WI only

Voluntary or Mandated:

Voluntary: Firm initiated