Exactech, Inc.: Medical Device Recall in 2024 - (Recall #: Z-1408-2024)
See the recall detail below. You can also see other recalls from the same firm in 2024.
Exactech Equinoxe GLENOID,POSTERIOR AUGMENT, PEGGED, 12o, LEFT, FOR CEMENTED USE ONLY, Mates with: a) 38, 41, 44, 47 head, Small, Item Number 314-04-22, b) 41, 44, 47, 50 head, Medium, Item Number 314-04-23, c) 44, 47, 50, 53 head, Large, Item Number 314-04-24, d) 44, 47, 50, 53 head, Extra Large, Item Number 314-04-25; Shoulder Arthroplasty
Class II
The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH).
a) Item Number 314-04-22, UDI/DI 10885862199553, Serial Numbers: 2111946, 2136867, 2136870, 2427405; b) Item Number 314-04-23, UDI/DI 10885862199560, Serial Numbers: 2111953, 2111954, 2139294, 2111955, 2139292, 2139291, 2139288, 2139290, 2424307, 2424304, 2111952, 2424306, 2424314, 2424305, 2482942; c) Item Number 314-04-24, UDI/DI 10885862199577, Serial Numbers: 2111949, 2111950, 2153613, 2153608, 2153607, 2153611, 2153612, 2421997, 2153609, 2421992, 2111951, 2153610, 2421994; d) Item Number 314-04-25, UDI/DI 10885862199584, Serial Numbers: 2112018, 2112019, 2427567
Nationwide and Australia, Austria, Belgium, Brazil, Chile, Canada , China, Colombia, France, Germany, Great Britain, Greece, Guatemala, India, Italy, Japan, Korea, Luxembourg, Malaysia, Netherlands, New Zealand, Portugal, Singapore, South Africa, Spain, Sweden, Switzerland, Thailand
Voluntary: Firm initiated