Excelsior Medical Corp: Medical Device Recall in 2014 - (Recall #: Z-2070-2014)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2014.

Data Source: FDA.
Product Description:

Excelsior Medical Sterile Triple Lead Tubing Set and Sterile Heavy Duty Triple Lead Tubing Set. To be used with the Pharm-Assist Dispensing Pump. Used for drug reconstitution and/or fluid transfer

Product Classification:

Class II

Date Initiated: June 5, 2014
Date Posted: July 23, 2014
Recall Number: Z-2070-2014
Event ID: 68573
Reason for Recall:

There is a potential for a leak. A molding defect in the connector that joins the three leads to the main flow line may allow fluid to escape or air to enter the fluid path during transfer of fluids from one container to another.

Status: Terminated
Product Quantity: 4250 tubing sets (PA03 - 78 cases; BX03 - 92 cases (each case contains 25 tubing sets))
Code Information:

Product Code PA03 - Lot Nos.: 03030813, 03050813 and 03091213 Product Code BX03, Lot Nos.: 64031513, 64073113, 64081313, 64090513

Distribution Pattern:

Worldwide Distribution - USA (nationwide) and Canada.

Voluntary or Mandated:

Voluntary: Firm initiated