Excelsior Medical Corp: Medical Device Recall in 2014 - (Recall #: Z-2070-2014)
See the recall detail below. You can also see other recalls from the same firm in 2014.
Excelsior Medical Sterile Triple Lead Tubing Set and Sterile Heavy Duty Triple Lead Tubing Set. To be used with the Pharm-Assist Dispensing Pump. Used for drug reconstitution and/or fluid transfer
Class II
There is a potential for a leak. A molding defect in the connector that joins the three leads to the main flow line may allow fluid to escape or air to enter the fluid path during transfer of fluids from one container to another.
Product Code PA03 - Lot Nos.: 03030813, 03050813 and 03091213 Product Code BX03, Lot Nos.: 64031513, 64073113, 64081313, 64090513
Worldwide Distribution - USA (nationwide) and Canada.
Voluntary: Firm initiated