Excelsior Medical Corp: Medical Device Recall in 2016 - (Recall #: Z-1989-2016)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2016.
Data Source: FDA.
Product Description:
Excelsior Disposable Syringe Heparin Lock Flush, USP syringe (5 mL Fill in 10 mL syringe). For flushing of IV catheters and IV tubing only.
Product Classification:
Class II
Date Initiated: April 28, 2016
Date Posted: June 22, 2016
Recall Number: Z-1989-2016
Event ID: 74045
Reason for Recall:
Certain lots of Heparin Lock Flush, USP Syringes (5 ml Fill in 10 ml Syringe) failed Anti-Factor IIa assay for potency during testing of retention samples for heparin potency.
Status: Terminated
Product Quantity: 184,500
Code Information:
Product Code, Lot number, Expiration date. 513604 3123326 7/1/2016 E0500-55 3123349 7/1/2016 513602 3126577 8/1/2017
Distribution Pattern:
US Nationwide distribution.
Voluntary or Mandated:
Voluntary: Firm initiated