Extremity Medical LLC: Medical Device Recall in 2013 - (Recall #: Z-0752-2013)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2013.

Data Source: FDA.
Product Description:

Extremity Medical, IO FiX System Accessory. Catalog Number 101-00010. This screw measuring device is an accessory to the IO FiX Screw and Washer System that is used to measure the length of the screw by placing the depth gauge over the guidewire and down to the bone. Orthopedic device.

Product Classification:

Class II

Date Initiated: March 22, 2012
Date Posted: February 6, 2013
Recall Number: Z-0752-2013
Event ID: 63805
Reason for Recall:

Lag screw measurement is short (approximately 5 mm) when using current screw measuring device (p/n 101-00010) per the IO FIX Surgical Technique Rev C. which can lead to possible inappropriate implant size selection (e.g. shorter screw size selected).

Status: Terminated
Product Quantity: 93 units
Code Information:

Catalog Number 101-00010 Lot number AP3V12 AP4E01 AP4N04 AP5E25 AP6A44 AP6E42 AP6L14 AP6M17 AP6T08

Distribution Pattern:

Worldwide Distribution -- Nationwide Distribution including the states of AR, AZ, CA, FL, GA, IL, IN, MD, MT, NC, NE, NH, NJ, NY, OH, PA, TN, TX, WI and WA.

Voluntary or Mandated:

Voluntary: Firm initiated