EYE COMFORT CARE LLC: Medical Device Recall in 2024 - (Recall #: Z-2586-2024)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2024.

Data Source: FDA.
Product Description:

Brand Name: UNclog Eye Mask Product Name: UNclog Eye Mask Model/Catalog Number: AG23JW24 Serial # ag22 Product Description: Heated eye mask Component: Heating unit, wire

Product Classification:

Class II

Date Initiated: February 8, 2024
Date Posted: August 21, 2024
Recall Number: Z-2586-2024
Event ID: 94929
Reason for Recall:

Due to a change in the safety clips for the heated eye masks, the wires to the safety clips may detach as a result of tugging and pulling and may cause a short in the heating element of the mask and could potentially cause burn injuries.

Status: Completed
Product Quantity: 500
Code Information:

Lot Code: AG23JW24 Serial # ag22

Distribution Pattern:

U.S. Nationwide distribution in the states of AL, AZ, CA, CT, FL, KS, KY, MA, ME, MT, NC, ND, NJ, NM, NY, OH, OR, PA, SC, TX, VA, WA, and WI.

Voluntary or Mandated:

Voluntary: Firm initiated