Fenwal Inc: Medical Device Recall in 2021 - (Recall #: Z-1345-2021)
See the recall detail below. You can also see other recalls from the same firm in 2021.
Amicus Exchange Kit Therapeutics - Product Usage: intended for use in the collection of blood components and mononuclear cells.
Class II
Some lots of Amicus MNC Apheresis and Amicus Exchange kits have leaking centrifuge packs during a procedure on the Amicus Separator using therapeutic protocols only. The defect is a blood leak at the boot, elbow, or in the channel on the separation chamber of the centrifuge pack.
Product Code X6R2349; UDI: 04086000101950; Batch Numbers FA20E25150 (Exp. 05/31/2022) and FA20I22142 (Exp. 09/30/2022)
US Nationwide distribution in the states of GA, TX, TN, MD, CA, KS, MN, MA, NM, NC, LA, CO, KY, NV, PA, UT, MI, NJ, IL, MD, OH, NY, FL.
Voluntary: Firm initiated