Ferno-Washington Inc: Medical Device Recall in 2015 - (Recall #: Z-2461-2015)
See the recall detail below. You can also see other recalls from the same firm in 2015.
Wheel caster set screw for Ferno iNX Wheeled Stretcher. Ferno iNX Wheeled Stretcher for pre-hospital patient transport.
Class II
The firm discovered an improperly installed set screw in one caster arm assembly on a unit in production. The screw assists in preventing the bearing cup on the caster assembly from backing out of the caster arm during use.
Serial numbers: 15F001703, 15F001704, 15F001705, 15F001706, 15F001707, 15F001708, 15F001709, 15F001710, 15F001711, 15F001712, 15F001713, 15F001714, 15F001715, 15F001716, 15F001717, 15F001719, 15F001720, 15F001721, 15F001722, 15F001723, 15F001724, 15F001725, 15F001726, 15F001727, 15F001729 & 15F001730.
Distributed in the states of AZ, CO, IA, MD, OH, PA & WA, and the countries of Australia, Canada & United Kingdom.
Voluntary: Firm initiated