Fisher Diagnostics: Medical Device Recall in 2017 - (Recall #: Z-2763-2017)
See the recall detail below. You can also see other recalls from the same firm in 2017.
Pacific Hemostasis FDP Latex Reagent, REF/Model Number 100651, UDI 00845275000870, 1 x 5 mL, IVD; --- Fisher Diagnostics, a division of Fisher Scientific Company, LLC, a part of Thermo Fisher Scientific Inc., Middletown, VA 22645-1905 USA
Class II
An issue was discovered with the Pacific Hemostasis Fibrinogen Degradation Products (FDP) kit, list number 100650, lot 948546 and FDP Latex Reagent, list 100651, lot 890199. While the kit performs accurately at FDP concentrations greater than 20 ¿g/ml, there is an observed reduction in sensitivity when FDP levels are less than 20 ¿g/ml. This reduction in sensitivity has been tied to FDP Latex Reagent lot 890199 and may result in false negatives.
Lot Number 890199, Exp. 6/30/2018
USA (nationwide) including Puerto Rico, and Internationally to Colombia
Voluntary: Firm initiated