Fisher & Paykel Healthcare, Ltd.: Medical Device Recall in 2024 - (Recall #: Z-1823-2024)
See the recall detail below. You can also see other recalls from the same firm in 2024.
Airvo 2 Humidifier, REF: PT101US, and myAirvo 2 Humidifier, REF: PT100US
Class II
Humidifier devices, used to deliver high flow respiratory therapy to patients, manufactured before 14 August 2017, have a speaker configuration that may result in distorted, intermittent or inaudible alarm sound levels. In the absence of an audible alarm, if there is an interruption to therapy, a patient may experience oxygen desaturation.
REF/UDI-DI: PT100US/09420012422248, PT101US 09420012422347. Serial Number Range: 120521XXXXXX - 170813XXXXXX Note: X digits are variable and do not affect the identification of the affected range. Devices manufactured before 14 August 2017
US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY.
Voluntary: Firm initiated