Fort Defiance Industries, LLC: Medical Device Recall in 2021 - (Recall #: Z-0121-2022)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2021.
Data Source: FDA.
Product Description:
P2131 Automated Field Steam Sterilizer NSN: 6530-01-641-4641
Product Classification:
Class II
Date Initiated: August 19, 2021
Date Posted: October 27, 2021
Recall Number: Z-0121-2022
Event ID: 88628
Reason for Recall:
The rear heater plate and cover assembly that contains the electrical wiring for the heaters does not have sufficient bonding to ground.
Status: Ongoing
Product Quantity: 380 devices
Code Information:
Model number: P2131; NSN: 6530-01-641-4641; Serial Numbers AFS-0001 through AFS-0386. Note: AFS-0297 was not manufactured.
Distribution Pattern:
Distributed nationwide to PA, MD, TX, NJ, OH, TN, UT, VA and internationally to Canada.
Voluntary or Mandated:
Voluntary: Firm initiated