Fresenius Medical Care Holdings, Inc. dba Renal Solutions: Medical Device Recall in 2013 - (Recall #: Z-2082-2013)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2013.

Data Source: FDA.
Product Description:

2008 MeDS Pump, Venofer(R) Pump Product Usage: The 2008 MeDs pump is an optional accessory for use with the Fresenius 2008 Series Hemodialysis Machines and is intended to administer Venofer to treat iron deficiency anemia in patients with chronic kidney disease undergoing hemodialysis, where Venofer, in conjunction with hemodialysis, is prescribed by a physician.

Product Classification:

Class II

Date Initiated: April 1, 2013
Date Posted: September 4, 2013
Recall Number: Z-2082-2013
Event ID: 65080
Reason for Recall:

Product is being recalled because the combination of the current 50 mg vial, the 2008 MeDS pump, and the 2008 MeDS pump tubing Set for CombiSet Bloodlines can lead to a situation where Venofer(R) delivery does not meet dosage accuracy requirements. This may lead to a slight over or under delivery of Venofer(R) that does not pose an acceptable risk to the patient. Please note that this recall

Status: Terminated
Product Quantity: 132
Code Information:

Serial numbers - TM-001065, TM-001283, TM-001284, TM-001325, TM-001411, TM-001426, TM-001465, TM-001476, TM-001529, TM-001967, TM-001968, TM-001975, TM-001985, TM-002015, TM-002016, TM-002017, TM-002018, TM-002021, TM-002022, TM-002095, TM-000015, TM-000028, TM-000031, TM-000041, TM-0048, TM-000050, TM-000051, TM-000053, TM-000055, TM-000059, TM-000064, TM-000065, TM-000068, TM-0069, TM-000213, TM-000215, TM-000218, TM-000219, TM-000070, TM-000072, TM-000074, TM-000075, TM-000078, TM-000081, TM-000082, TM-000083, TM-000085, TM-000086, TM-000087, TM-000088, TM-000089, TM-000092, TM-000040, TM-000077, TM-000098, TM-000101, TM-000102, TM-000103, TM-000104, TM-000105, TM-000106, TM-000107, TM-000108, TM-000109, TM-000112, TM-000113, TM-000114, TM-000115, TM-000116, TM-000122, TM-000123, TM-000124, TM-000125, TM-000126, TM-000128, TM-000145, TM-000002, TM-000003, TM-000004, TM-000005, TM-000006, TM-000007, TM-000008, TM-000009, TM-000010, TM-000012, TM-000013, TM-000014, TM-000017, TM-000018, TM-000019, TM-000020, TM-000023, TM-000024, TM-000025, TM-000033, TM-000036, TM-A595, TM-A610, TM-A627, TM-000220, TM-000227, TM-007364, TM-007371, TM-006848, TM-006860, TM-006875, TM-006879, TM-006880, TM-006881, TM-006882, TM-006883, TM-006884, TM-006885, TM-006886, TM-006887, TM-006890, TM-006891, TM-006892, TM-006893, TM-006897, TM-006898, TM-006911, TM-006912, TM-006913, TM-006914, TM-006915, TM-006916, TM-006917, TM-006918, TM-006919, and TM-006920.

Distribution Pattern:

USA Nationwide Distribution in the states of CA, MA, and NY.

Voluntary or Mandated:

Voluntary: Firm initiated