Product Description:

2008 Series Hemodialysis Machines: 2008K2 with the following product code and description: 190633 2008K2 HEMO SYSTEM, OPTIONAL MODULE; 190618 2008K2 HEMODIALYSIS SYS OLC/DP, SPANISH; 190610 2008K2 HEMODIALYSIS SYS. OLC/DIASAFE PLS; 190630 2008K2 MACHINE, SHORT CAB, OLC/DP, HP; 190628 2008K2 OLC ONLY WITH HEPARIN PUMP The Fresenius 2008K is indicated for acute and chronic dialysis therapy.

Product Classification:

Class II

Date Initiated: February 20, 2015
Date Posted: May 6, 2015
Recall Number: Z-1552-2015
Event ID: 70574
Reason for Recall:

The Acetate value for GranuFlo on the Select Concentrate screen should be 8mEq/l but it is displayed as 4mEq/l. The correct value is shown on the dialysate screen.

Status: Terminated
Product Quantity: 133,751 in total (US: 130,337 and OUS: Canada 1,201 and Mexico 2,213)
Code Information:

Serial Numbers: 0K0S-0678 0K0S-0885 0K0S-0901 0K0S-0917 0K0S-0922 0K0S-0927 0K0S-0935 0K0S-0943 0K0S-0959 0K0S-0960 0K0S-0963 0K0S-0992 0K0S149209-149999 0K0S150000- 152000 0K0S152001- 154000 0K0S154001- 160000 0K0S160001- 162000 0K0S162001- 165353 0S154030 1K0S-1000 1K0S-1022 1K0S-1027 1K0S-1029 1K0S-1035 1K0S-1037 1K0S-1044- 1046 1K0S-1065 1K0S-1067 1K0S-1070 1K0S-1094 1K0S-1095 1K0S-1102 1K0S-1103 1K0S-1121 1K0S165354- 174103 2K0S174104 2K0S174105- 174111 2K0S174150- 182134 2K0S-5783 2K0S-5802 2K0S-6015 2K0S-6046 2K0S-6065 2K0S-6067 2K0S-6109 2K0S-6232 2K0S-6266 2K0S-9573 3K0S182135- 188639 4K0S188640- 192508 4K0S192542 4K0S192545 4K0S192550- 192553 4K0S192555 4K0S192566- 192569 4K0S192571 4K0S192572 4K0S192575- 192577 4K0S192579 4K0S192581 4K0S192584- 192586 4K0S192588 4K0S192593 4K0S192602 4K0S192603 4K0S192605 4K0S192675 4K0S192738- 192740 4K0S-N592 4K0S-Q302 4K0S-Q350 4K0S-Q362 4K0S-Q425 4K0S-Q467 4K0S-Q470 4K0S-Q502 5K0S-132B 5K0S-856B 5K0S-T230 5K0S-V420 5K0S-X901 5K0S-Y378 6K0S-029N 6K0S-151Q 6K0S-180Q 6K0S-205M 6K0S-239M 6K0S-255N 6K0S-315N 6K0S-331Q 6K0S-821G 6K0S-903R 6K0S-942S 7K0S103279 7K0S103983 7K0S104826 7K0S112150 7K0S112155 7K0S115550 8K0S118640 8K0S121287 8K0S122793 8K0S126296- 126305 8K0S126314- 126323 8K0S126573 8K0S128448 8K0S128506 8K0S129030- 129033 8K0S129211- 129321 8K0S129335 8K0S129337- 129339 8K0S129369 8K0S129431- 129435 8K0S129440 8K0S129447- 129490 8K0S129503- 129516 8K0S129893 8K0S130289 8K0S130296 8K0S130300 8K0S130304 8K0S130324 8K0S130345 8K0S130347 8K0S130546 8K0S130548 8K0S130552 8K0S130556 8K0S130559 8K0S130565 8K0S130567 8K0S130571 8K0S130574 8K0S130726 8K0S130751 8K0S130768 8K0S130771- 130773 8K0S130819 8K0S130951 8K0S131043 8K0S131046 8K0S131120 8K0S131133 8K0S131328 8K0S131330 8K0S131449- 132162 9K0S132163- 135000 9K0S135001- 149208 9KOS137160 OKOS149422

Distribution Pattern:

Worldwide Distribution: US (Nationwide) and countries of: Canada and Mexico.

Voluntary or Mandated:

Voluntary: Firm initiated