Fresenius Medical Care Holdings, Inc.: Medical Device Recall in 2021 - (Recall #: Z-0834-2021)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2021.
Data Source: FDA.
Product Description:
Citrasate Liquid Acid concentrate bottles, Product Code 08-2251-CA - Product Usage: used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine.
Product Classification:
Class II
Date Initiated: November 23, 2020
Date Posted: January 20, 2021
Recall Number: Z-0834-2021
Event ID: 86939
Reason for Recall:
The deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously communicated. Calcium concentration is below manufacturing specification.
Status: Ongoing
Product Quantity: 278 cases
Code Information:
Lot Numbers: 20NXAC007 Additional Affected lots as of 3/17/21: 20NXAC007
Distribution Pattern:
US Nationwide distribution.
Voluntary or Mandated:
Voluntary: Firm initiated