Fresenius Medical Care Holdings, Inc.: Medical Device Recall in 2021 - (Recall #: Z-0838-2021)
See the recall detail below. You can also see other recalls from the same firm in 2021.
Naturalyte Liquid Acid concentrate bottles, Product Code 08-4231-0 - Product Usage: used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine.
Class II
The deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously communicated. Calcium concentration is below manufacturing specification.
Lot Numbers: 20LXAC088 Additional Affected lots as of 3/17/21: 20LXAC088 20LXAC052
US Nationwide distribution.
Voluntary: Firm initiated