Fresenius Medical Care Holdings, Inc.: Medical Device Recall in 2021 - (Recall #: Z-0841-2021)
See the recall detail below. You can also see other recalls from the same firm in 2021.
Citrasate Liquid Acid concentrate drums, Product Code 13-2251 -CA - Product Usage: used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine.
Class II
The deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously communicated. Calcium concentration is below manufacturing specification.
Lot Numbers: 20LXAC033 20NXAC004 Additional Affected lots as of 3/17/21: 20LXAC031 20NXAC004 20LXAC033 20NXAC003 20LXAC032
US Nationwide distribution.
Voluntary: Firm initiated