Fresenius Medical Care Holdings, Inc.: Medical Device Recall in 2021 - (Recall #: Z-0841-2021)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2021.

Data Source: FDA.
Product Description:

Citrasate Liquid Acid concentrate drums, Product Code 13-2251 -CA - Product Usage: used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine.

Product Classification:

Class II

Date Initiated: November 23, 2020
Date Posted: January 20, 2021
Recall Number: Z-0841-2021
Event ID: 86939
Reason for Recall:

The deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously communicated. Calcium concentration is below manufacturing specification.

Status: Ongoing
Product Quantity: 2 drums
Code Information:

Lot Numbers: 20LXAC033 20NXAC004 Additional Affected lots as of 3/17/21: 20LXAC031 20NXAC004 20LXAC033 20NXAC003 20LXAC032

Distribution Pattern:

US Nationwide distribution.

Voluntary or Mandated:

Voluntary: Firm initiated