Fresenius Medical Care Holdings, Inc.: Medical Device Recall in 2022 - (Recall #: Z-0734-2023)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2022.

Data Source: FDA.
Product Description:

2008T and 2008TBluestar Hemodialysis Machines with CDX-Indicated for acute and chronic dialysis therapy in a healthcare facility Models: (1) 190713 2008T HEMODIALYSIS SYS., with CDX; (2) 190766 2008T HEMODIALYSIS SYSTEM W/BIBAG W/CDX ; (3) 191124 2008T HD SYS. CDX BLUESTAR; (4) 191126 2008T HD SYS. CDX W/BIBAG BLUESTAR

Product Classification:

Class II

Date Initiated: July 27, 2022
Date Posted: December 28, 2022
Recall Number: Z-0734-2023
Event ID: 91154
Reason for Recall:

(1)Software version 2.74 upgrade and (2)hardware related cause that affects only CDX Machines, has been identified as the UI-MICs Board is sensitive to Electromagnetic Interference (EMI) which will cause the false alarm "Remove USB 2 Device".

Status: Ongoing
Product Quantity: 2008T with CDX= 70,552 2008T Bluestar with CDX= 25,205
Code Information:

Software issue in SW versions 2.72 and earlier with CDX. UDI: (1) 00840861100897; (2) 00840861100910; (3) 00840861102082; (4) 00840861102099

Distribution Pattern:

US Nationwide distribution.

Voluntary or Mandated:

Voluntary: Firm initiated