Fresenius Medical Care Holdings, Inc.: Medical Device Recall in 2025 - (Recall #: Z-0936-2026)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2025.
Data Source: FDA.
Product Description:
5008X CAREsystem +CLiC +CDX;
Product Classification:
Class II
Date Initiated: October 25, 2025
Date Posted: December 24, 2025
Recall Number: Z-0936-2026
Event ID: 97934
Reason for Recall:
Several software anomalies with the potential to impact patient treatment, of which one includes the possibility to result in treatment stop if certain conditions are fulfilled.
Status: Ongoing
Product Quantity: 115 units
Code Information:
Model Number: M204441; UDI-DI: 00840861102433; All serial numbers produced with a software version prior to 4.82.4;
Distribution Pattern:
Domestic: LA, MA, MN;
Voluntary or Mandated:
Voluntary: Firm initiated