Fresenius Medical Care Renal Therapies Group, LLC: Medical Device Recall in 2015 - (Recall #: Z-0294-2016)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2015.
Data Source: FDA.
Product Description:
Fresenius 132 Gallon GranuFlo Dissolution Unit - P/N G047-80101. For Use in the Preparation of acid concentrate for hemodialysis.
Product Classification:
Class II
Date Initiated: October 23, 2015
Date Posted: November 25, 2015
Recall Number: Z-0294-2016
Event ID: 72484
Reason for Recall:
Dissolution units and replacement motors units do not provide full thermal protection for both the Main and Starter coils resulting in the generation of smoke and/or potentially lead to fire.
Status: Terminated
Product Quantity: 918 units
Code Information:
All 132 Gallon Granuflo Dissolution Units and 132 Gallon Dry Acid Dissolution Units with the H218 Motor and Modified H218 Mixer Motors.
Distribution Pattern:
Nationwide Distribution.
Voluntary or Mandated:
Voluntary: Firm initiated