Fresenius Medical Care Renal Therapies Group, LLC: Medical Device Recall in 2016 - (Recall #: Z-0693-2016)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2016.
Data Source: FDA.
Product Description:
Fresenius Liberty Cycler-Peritoneal Dialysis Cycler Model: 180111 and RTLR180111
Product Classification:
Class II
Date Initiated: December 15, 2015
Date Posted: February 3, 2016
Recall Number: Z-0693-2016
Event ID: 72926
Reason for Recall:
The door latch may not fully close and the door may open unexpectedly. The door may make physical contact with anything near the door of the machine, including the patient. May cause injury, delay in treatment, or Increased Intraperitoneal Volume (IIPV) (overfill).
Status: Terminated
Product Quantity: 14,094 units
Code Information:
All lot numbers
Distribution Pattern:
Nationwide Distribution
Voluntary or Mandated:
Voluntary: Firm initiated