Fresenius Medical Care Renal Therapies Group, LLC: Medical Device Recall in 2016 - (Recall #: Z-0887-2016)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2016.
Data Source: FDA.
Product Description:
Fresenius Naturalyte Liquid Acid Concentrate Catalog Number: 08-2301-3 Intended Use: Acid Concentrate for Bicarbonate Dialysis
Product Classification:
Class II
Date Initiated: January 18, 2016
Date Posted: March 2, 2016
Recall Number: Z-0887-2016
Event ID: 73077
Reason for Recall:
On the label calcium (Ca) listed in the right upper corner of the label (red background) incorrectly indicates the product contains 2.5 mEq/L Ca. The concentration indicated in the list of constituents represents the actual calcium concentration of the final dialysate, 3.0 mEq/L.
Status: Terminated
Product Quantity: 848 (212 cases)
Code Information:
Lot Number: 15STAC072 Exp Date: 2017-12-31
Distribution Pattern:
US Distribution to states of: KS, LA, MS, OK, and TX.
Voluntary or Mandated:
Voluntary: Firm initiated