Fresenius Medical Care Renal Therapies Group, LLC: Medical Device Recall in 2017 - (Recall #: Z-1026-2017)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2017.

Data Source: FDA.
Product Description:

Fresenius 2008T Series Hemodialysis System

Product Classification:

Class II

Date Initiated: December 21, 2016
Date Posted: January 25, 2017
Recall Number: Z-1026-2017
Event ID: 76037
Reason for Recall:

When the UF Rate, Goal or Time is adjusted using the up and down arrow keys, and the change is cancelled by using the esc key, the cancelled UF Rate is actually being executed rather than rate displayed on the machine

Status: Terminated
Product Quantity: 65361 units (US) 249 units (OUS)
Code Information:

All serial numbers

Distribution Pattern:

Nationwide Foreign: Canada, Mexico, Jamaica, Germany, Guam

Voluntary or Mandated:

Voluntary: Firm initiated