Fresenius Medical Care Renal Therapies Group, LLC: Medical Device Recall in 2018 - (Recall #: Z-1010-2018)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2018.
Data Source: FDA.
Product Description:
NaturaLyte¿ Liquid Acid Concentrate 2.0 mEq/L, Model # 08-2251-0 3.43 liter bottle This concentrate is formulated to be used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine.
Product Classification:
Class II
Date Initiated: September 29, 2017
Date Posted: March 21, 2018
Recall Number: Z-1010-2018
Event ID: 79236
Reason for Recall:
The label specifies the potassium concentration is 2.0 mEq/L. Some of the product in the affected lot was found to have a potassium concentration of 2.25 mEq/L.
Status: Terminated
Product Quantity: 598 cases
Code Information:
Lot # 17KTAC076; UDI-00840861101696
Distribution Pattern:
US Distribution to states of: CA, FL, KS, LA, and TX.
Voluntary or Mandated:
Voluntary: Firm initiated