FujiFilm Healthcare Americas Corporation: Medical Device Recall in 2021 - (Recall #: Z-0189-2022)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2021.

Data Source: FDA.
Product Description:

Arietta 750 Ultrasound Software Version: V1.0.0 through V2.1.3

Product Classification:

Class II

Date Initiated: October 14, 2021
Date Posted: November 10, 2021
Recall Number: Z-0189-2022
Event ID: 88840
Reason for Recall:

Diagnostic ultrasound system with the specified software version and used in conjunction with a Fujifilm (formally Hitachi) transducer does not display the measurement results correctly with two cardiology measurement functions: Mitral Regurgitation Flow (MR Flow) Measurement; Mitral Valve (MV) Measurement. In addition there is an error in the instruction manual.

Status: Ongoing
Product Quantity: 15 units
Code Information:

Serial Numbers: G3075073 (01)04562122959841(11)201214(21)G3075073; G3110777 (01)04562122959841(11)210421(21)G3110777; G3110771 (01)04562122959841(11)210421(21)G3110771; G3113780 (01)04562122959841(11)210510(21)G3113780; G3113778 (01)04562122959841(11)210510(21)G3113778; G3113779 (01)04562122959841(11)210510(21)G3113779; G3113777 (01)04562122959841(11)210510(21)G3113777; G3071857 (01)04562122959841(11)210420(21)G3071857; G3108384 (01)04562122959841(11)210402(21)G3108384; G3114220 (01)04562122959841(11)210510(21)G3114220; G3110776 (01)04562122959841(11)210421(21)G3110776; G3108385 (01)04562122959841(11)210402(21)G3108385; G3114223 (01)04562122959841(11)210518(21)G3114223; G3108401 (01)04562122959841(11)210420(21)G3108401; G3114959 (01)04562122959841(11)210524(21)G3114959

Distribution Pattern:

Nationwide

Voluntary or Mandated:

Voluntary: Firm initiated