FUJIFILM Healthcare Americas Corporation: Medical Device Recall in 2022 - (Recall #: Z-1750-2022)
See the recall detail below. You can also see other recalls from the same firm in 2022.
Surpria 64: Software Version V3.11, V3.22
Class II
Error 00003050 or 00003052 may occur during scanning, the captured images will not be reconstructed and the Raw Data will not be displayed in the work list.result, may result in the need for additional scanning to capture the images properly for reconstruction, thereby exposing the patient to additional radiation
UDI: (01)04573596200520(11)220215(21)W5086(250)G1Z0211021 (01)04580292766601(11)200130(21)W8013(250)G1Z0098019 (01)04580292766601(11)200226(21)W8014(250)G1Z0130719 (01)04580292766601(11)201217(21)W8015(250)G1Z0024020 (01)04580292766601(11)210217(21)W8016(250)G1Z0216720 (01)04580292766601(11)210528(21)W8017(250)G1Z0239620 (01)04580292766601(11)210610(21)W8018(250)G1Z0039321 (01)04580292766601(11)210628(21)W8019(250)G1Z0223420 (01)04573596200520(11)210727(21)W8020(250)G1Z0087821 (01)04573596200520(11)210825(21)W8021(250)G1Z0139021 (01)04573596200520(11)211012(21)W8022(250)G1Z0112721 (01)04573596200520(11)211026(21)W8023(250)G1Z0156921 (01)04573596200520(11)220113(21)W8024(250)G1Z0043121 (01)04573596200520(11)220113(21)W8025(250)G1Z0240221 (01)04573596200520(11)220204(21)W8026(250)G1Z0244021 (01)04573596200520(11)220204(21)W8027(250)G1Z0249821 (01)04573596200520(11)220214(21)W8028(250)G1Z0249921 Serial Numbers: G1Z0098019 G1Z0130719 G1Z0024020 G1Z0216720 G1Z0239620 G1Z0039321 G1Z0223420 G1Z0087821 G1Z0139021 G1Z0112721 G1Z0156921 G1Z0043121 G1Z0240221 G1Z0244021 G1Z0249821 G1Z0249921
US Nationwide Distribution
Voluntary: Firm initiated