FUJIFILM Healthcare Americas Corporation: Medical Device Recall in 2024 - (Recall #: Z-1884-2024)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2024.
Data Source: FDA.
Product Description:
Noblus AC Adapter used with the Noblus Ultrasound Imaging System Model : AHM250PS24
Product Classification:
Class II
Date Initiated: April 11, 2024
Date Posted: May 29, 2024
Recall Number: Z-1884-2024
Event ID: 94441
Reason for Recall:
AC adapters of Noblus can heat up, and cause smoke in the worst case, when an excessive mechanical stress is applied to the socket for power cable (frequent insertion/removal of the power cable).
Status: Ongoing
Product Quantity: 418 units
Code Information:
UDI-DI: (01) 04580292771704 Noblus Ultrasound Imaging System AC Adapter Serial Number Before K17160001 (Excluding K17160001)
Distribution Pattern:
US Nationwide distribution.
Voluntary or Mandated:
Voluntary: Firm initiated