Fujirebio Diagnostics, Inc.: Medical Device Recall in 2026 - (Recall #: Z-1303-2026)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2026.

Data Source: FDA.
Product Description:

Product Name: Lumipulse G pTau 217 Plasma Calibrators Model/Catalog Number: 81557 Software Version: Not Applicable Product Description: Lumipulse G pTau 217 Plasma Calibrators: The pTau 217 plasma concentration of a specimen is automatically calculated from the calibration curve, which is also automatically calculated from calibration data. The result of the calculation is reported in pg/mL. Lumipulse G pTau 217 Plasma Calibrators: Liquid (Frozen), 1x1.5 mL (5 Concentrations) CAL 1 0 pg/mL pTau 217 Plasma Calibrator CAL 2 0.250 pg/mL pTau 217 Plasma Calibrator CAL 3 1.000 pg/mL pTau 217 Plasma Calibrator CAL 4 5.000 pg/mL pTau 217 Plasma Calibrator CAL 5 10.000 pg/mL pTau 217 Plasma Calibrator Contains Tris buffer with protein (bovine) and chemical stabilizers. Preservative: 0.1% ProClin 300, 0.05% ProClin 950 Component: Yes. Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio

Product Classification:

Class II

Date Initiated: December 11, 2025
Date Posted: February 11, 2026
Recall Number: Z-1303-2026
Event ID: 98232
Reason for Recall:

Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low specificity when compared back to FDA-cleared CSF ratio/PET imaging, due to falsely elevated positive ratio or falsely elevated indeterminate ratio results causing an incorrect classification of a patient as having amyloid pathology associated with Alzheimer s disease (positive result) and/or requiring more clinical testing of that patient (indeterminate result).

Status: Ongoing
Product Quantity: 117 units
Code Information:

Model/Catalog Number: 81557; UDI-DI: 05414470815579; Lot Numbers: D5C6094U, D5C6125U;

Distribution Pattern:

US Nationwide distribution in the states of Arizona, California, Florida, Indiana, Kansas, Kentucky, Maryland, Massachusetts, Minnesota, New Jersey, Pennsylvania, Texas, Washington.

Voluntary or Mandated:

Voluntary: Firm initiated