Galt Medical Corporation: Medical Device Recall in 2020 - (Recall #: Z-0568-2021)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2020.

Data Source: FDA.
Product Description:

Centeze Centesis Catheter, 6F, 10cm, 18Ga, 12cm, REF DRC-002-06 Centeze Centesis Catheter, 6F, 10cm, 18Ga, 12cm, REF DRC-006-16

Product Classification:

Class II

Date Initiated: November 16, 2020
Date Posted: December 23, 2020
Recall Number: Z-0568-2021
Event ID: 86795
Reason for Recall:

Due to a manufacturing issue, the catheter tip geometry is incorrect (thicker wall) resulting in a more obtuse tip angle.

Status: Terminated
Product Quantity: 9,920 catheters (updated on 03/17/2021)
Code Information:

Catalog Number: DRC-002-06 Lot Numbers/UDI Code: 20108509/(01)0084 1268 100398(17)240428(10)20108509 20139676/(01)0084 1268 100398(11)200520(17)240519(10)20139676 20148911/(01)0084 1268 100398(11)200609(17)240608(10)20148911 20160253/(01)0084 1268 100398(11)200707(17)240706(10)20160253 20189100/(01)0084 1268 100398(11)200821(17)240820(10)20189100 Updated Information provided 03/16/2021: For Catalog Number: DRC-002-06 additional Lot Numbers: -20080351 -20234573 -20297230 -21008452 Additional Products: Catalog Number: DRC-006-16 Lot Numbers: -20217235 -20217236 -20217237

Distribution Pattern:

U.S.: FL, GA, MA, NC, NJ, OH, PA, Puerto Rico and UT. O.U.S: Canada, Netherlands, Turkey, United Kingdom, and UAE

Voluntary or Mandated:

Voluntary: Firm initiated