GE Healthcare, LLC: Medical Device Recall in 2018 - (Recall #: Z-0608-2019)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2018.
Data Source: FDA.
Product Description:
Discovery NM/CT 670 Pro, model 5376204-70-54
Product Classification:
Class II
Date Initiated: May 24, 2018
Date Posted: December 19, 2018
Recall Number: Z-0608-2019
Event ID: 81680
Reason for Recall:
When your system was installed, the Main disconnect switch, EPO button on the A1 panel, and Remote EPO button(s) (outside of the A1 panel) may not have been wired correctly which could result in these not shutting off all power to the entire system as intended, resulting in a potential electric hazard.
Status: Completed
Product Quantity: 1 unit
Code Information:
GTIN 00840682121194 Serial Number ESDX35015
Distribution Pattern:
CA, DC, OH, UT, WI, Canada
Voluntary or Mandated:
Voluntary: Firm initiated