GE Healthcare, LLC: Medical Device Recall in 2018 - (Recall #: Z-2583-2018)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2018.

Data Source: FDA.
Product Description:

CT Goldseal Optima CT 600

Product Classification:

Class II

Date Initiated: May 24, 2018
Date Posted: August 8, 2018
Recall Number: Z-2583-2018
Event ID: 80484
Reason for Recall:

Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).

Status: Terminated
Product Quantity: 42 units total
Code Information:

Serial Number 51941YC3, System ID Number 406434660

Distribution Pattern:

Nationwide distribution to AL, CA, CO, FL, IL, IN, MA, MD, MI, MS, NC, NJ, NY,OH, OK, SC, TN, TX, VA.

Voluntary or Mandated:

Voluntary: Firm initiated