GE Healthcare, LLC: Medical Device Recall in 2018 - (Recall #: Z-2586-2018)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2018.
Data Source: FDA.
Product Description:
Optima CT 540
Product Classification:
Class II
Date Initiated: May 24, 2018
Date Posted: August 8, 2018
Recall Number: Z-2586-2018
Event ID: 80484
Reason for Recall:
Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).
Status: Terminated
Product Quantity: 42 units total
Code Information:
Serial number 500925HM3, System ID Number 615320OPT
Distribution Pattern:
Nationwide distribution to AL, CA, CO, FL, IL, IN, MA, MD, MI, MS, NC, NJ, NY,OH, OK, SC, TN, TX, VA.
Voluntary or Mandated:
Voluntary: Firm initiated