GE Healthcare, LLC: Medical Device Recall in 2020 - (Recall #: Z-0819-2020)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2020.

Data Source: FDA.
Product Description:

Centricity Universal Viewer 6.0, a device that displays medical images and data from various imaging sources, and from other healthcare information sources.

Product Classification:

Class II

Date Initiated: December 16, 2019
Date Posted: January 29, 2020
Recall Number: Z-0819-2020
Event ID: 84584
Reason for Recall:

Centricity Universal Viewer measurements saved into a DICOM Grayscale Presentation State are incorrect in subsequent views for exams containing series with different pixel sizes and may lead to a potential misdiagnosis.

Status: Terminated
Product Quantity: 205
Code Information:

Model 2088026-9XX Versions 6.0 SP10 or higher

Distribution Pattern:

Distributed nationwide; Internationally distributed to Mexico and 24 additional countries.

Voluntary or Mandated:

Voluntary: Firm initiated