GE Healthcare, LLC: Medical Device Recall in 2021 - (Recall #: Z-0242-2022)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2021.
Data Source: FDA.
Product Description:
Computed Tomography Systems Revolution CT, Revolution CT ES, Revolution Apex, Revolution CT with Apex edition
Product Classification:
Class II
Date Initiated: June 30, 2021
Date Posted: December 1, 2021
Recall Number: Z-0242-2022
Event ID: 89027
Reason for Recall:
The accumulated dose is incorrectly displayed in certain situations. Thisoccurs during an interventional exam using the SmartStep option if the table height is adjusted after the exam starts. In this situation, the SmartStep display will show a value that is higher (up to 100 times) than the actual dose delivered.
Status: Ongoing
Product Quantity: 105 in total
Code Information:
Revolution CT, Revolution CT ES, Revolution Apex, Revolution CT with Apex edition
Distribution Pattern:
Worldwide and US Nationwide Distribution
Voluntary or Mandated:
FDA Mandated