GE Medical Systems China Co., Ltd.: Medical Device Recall in 2025 - (Recall #: Z-1637-2025)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2025.

Data Source: FDA.
Product Description:

GE Healthcare CARESTATION 750c A1, Model/REF Number 1012-9755-000

Product Classification:

Class I

Date Initiated: March 21, 2025
Date Posted: May 7, 2025
Recall Number: Z-1637-2025
Event ID: 96634
Reason for Recall:

GE HealthCare has become aware that certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.

Status: Ongoing
Product Quantity: 74 units
Code Information:

GTIN 00840682146425, Serial Numbers: SQT23330001WA, SQT23080006WA, SQT23080007WA, SQT23080008WA, SQT23080009WA, SQT23080010WA, SQT23080011WA, SQT24440011WA, SQT23150001WA, SQT23070001WA, SQT23070002WA, SQT24430006WA, SQT23200001WA, SQT23200002WA, SQT23200003WA, SQT23200004WA, SQT23200005WA, SQT24290003WA, SQT24510015WA, SQT24520016WA, SQT23360008WA, SQT23360002WA, SQT23360003WA, SQT23360004WA, SQT23360005WA, SQT23360006WA, SQT23360007WA, SQT23170001WA, SQT24430007WA, SQT24430008WA, SQT24430009WA, SQT24430010WA, SQT23190004WA, SQT24410005WA, SQT23050001WA, SQT23050002WA, SQT23050003WA, SQT23050004WA, SQT23060001WA, SQT23060002WA, SQT23060003WA, SQT23300002WA, SQT24230002WA, SQT24440012WA, SQT24440013WA, SQT24440014WA, SQT23400001WA, SQT23400002WA, SQT24022000WA, SQT23480001WA, SQT23370002WA, SQT23370007WA, SQT23370004WA, SQT23370006WA, SQT23370005WA, SQT23370001WA, SQT23370003WA, SQT23370008WA, SQT23370009WA, SQT23370010WA, SQT23370011WA, SQT23460001WA, SQT24160001WA, SQT23350001WA, SQT23420001WA, SQT23420002WA, SQT23190001WA, SQT23190002WA, SQT23190003WA, SQT23190005WA, SQT23190006WA, SQT23190007WA, SQT23190008WA, SQT23190009WA.

Distribution Pattern:

Worldwide distribution. US nationwide including Puerto Rico, Albania, Algeria, Argentina, Armenia, Australia, Austria, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Fiji, Finland, France, Germany, Greece, Guatemala, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Republic of Korea, Kosovo, Kuwait, Latvia, Lebanon, Libya, Lithuania, Malaysia, Maldives, Malta, Mexico, Moldova, Montenegro, Morocco, Myanmar, Netherlands, New Zealand, Nigeria, North Macedonia, Norway, Oman, Pakistan, State of Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Rwanda, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tonga, Trinidad Tobago, Tunisia, T¿rkiye, Uganda, Ukraine, UAE, United Kingdom, Uruguay, Uzbekistan, Venezuela, and Vietnam.

Voluntary or Mandated:

Voluntary: Firm initiated