GE Medical Systems Information Technologies Inc: Medical Device Recall in 2023 - (Recall #: Z-2613-2023)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2023.

Data Source: FDA.
Product Description:

Prucka 3 Amplifiers, Model P1801PA, and Field Replaceable Units (FRUs), Model 5875569, used with CardioLab AltiX and ComboLab AltiX.

Product Classification:

Class II

Date Initiated: August 8, 2023
Date Posted: September 27, 2023
Recall Number: Z-2613-2023
Event ID: 92903
Reason for Recall:

A diode on the power supply of the Prucka 3 Amplifier used with CardioLab/ComboLab systems could reach elevated temperatures and fail, then power off and become inoperable.

Status: Ongoing
Product Quantity: 81 amplifiers
Code Information:

GTIN 00195278507044 (CardioLab) and 00195278507051 (ComboLab) Serial numbers for Amplifiers: SVS22330007SA, SVS22330008SA, SVS22330009SA, SVS22330010SA, SVS22330011SA, SVS22330012SA, SVS22330013SA, SVS22330014SA, SVS22330016SA, SVS22330017SA, SVS22330018SA, SVS22330019SA, SVS22330020SA, SVS22330021SA, SVS22330022SA, SVS22330023SA, SVS22330024SA, SVS22330025SA, SVS22330026SA, SVS22330028SA, SVS22330030SA, SVS22330031SA, SVS22330032SA, SVS22330033SA, SVS22330034SA, SVS22330035SA, SVS22440001SA, SVS22440002SA, SVS22440005SA, SVS22440008SA, SVS22440009SA, SVS22440010SA, SVS22440011SA, SVS22440012SA, SVS22440013SA, SVS22440016SA, SVS22490002SA, SVS22490003SA, SVS22490004SA, SVS22520006SA, SVS22520008SA, SVS22520009SA, SVS22520010SA, SVS22520011SA, SVS22520012SA, SVS22520013SA, SVS22520015SA, SVS23040001SA, SVS23040002SA, SVS23040003SA, SVS23040004SA, SVS23040005SA, SVS23040007SA, SVS23040008SA, SVS23040011SA, SVS23040012SA, SVS23040013SA, SVS23040014SA, SVS23040015SA, SVS23040016SA, SVS23040017SA, SVS23040018SA, SVS23040019SA, SVS23070002SA, SVS23070004SA, SVS23070005SA, SVS23070008SA, SVS23070009SA, SVS23070011SA, SVS23070012SA, SVS23100001SA, SVS23100002SA, SVS23100004SA, and SVS23100006SA. Serial numbers for FRUs: SVS22330015SA, SVS22440006SA, SVS22440007SA, SVS22440015SA, SVS22490001SA, SVS22520005SA, and SVS22520014SA.

Distribution Pattern:

Distribution was made to CA, CO, FL, GA, IL, IN, KY, MA, NC, NJ, OH, OR, PA, SC, TX, VA, and WI. There was government distribution but no military distribution. Foreign distribution was made to Canada, France, Germany, Greece, Ireland, Japan, Republic of Korea, and Romania.

Voluntary or Mandated:

Voluntary: Firm initiated