GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING: Medical Device Recall in 2023 - (Recall #: Z-1057-2023)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2023.

Data Source: FDA.
Product Description:

GE Nuclear Medicine 600/800 series systems, specifically: a) 870 CZT O540+ Gantry 7.25mm, b) D670 CZT Gantry 7.25mm, c) DISCOV NM/CT 870 CZT, d) Discovery NMCT 670 CZT O540+, e) NM/CT 870 CZT CYBER DEFENCE, f) NM/CT 870 CZT O540, g) NM/CT 870 CZT WSO, h) NM/CT 870 CZT, i) NM/CT 870 CZT O540+ 7.25MM; Emission Computed Tomography System

Product Classification:

Class I

Date Initiated: December 19, 2022
Date Posted: February 15, 2023
Recall Number: Z-1057-2023
Event ID: 91456
Reason for Recall:

A mitigation may not be correctly implemented. If this is the case, the detector can fall, leading to life-threatening bodily injury.

Status: Ongoing
Product Quantity: 86 units
Code Information:

a) 870 CZT O540+ Gantry 7.25mm: Catalog number H3906CW , UDI/DI 01008406821408432187CZ6304211200300, Serial Numbers: 87CZ63042; UDI/DI 01008406821408432187CA6305311210100, Serial Numbers: 87CA63053; UDI/DI 01008406821408432187CX6301111181000, Serial Numbers: 87CX63011; UDI/DI 01008406821408432187CY6302011190300, Serial Numbers: 87CY63020; UDI/DI 01008406821408432187CY6303511191100, Serial Numbers: 87CY63035; UDI/DI 01008406821408432187CZ6304011200100, Serial Numbers: 87CZ63040; UDI/DI 01008406821408432187CX6300411180700, Serial Numbers: 87CX63004; UDI/DI 01008406821408432187CX6300911180900, Serial Numbers: 87CX63009; UDI/DI 01008406821408432187CB6307211220700, Serial Numbers: 87CB63072; UDI/DI 01008406821408432187CA6306111211100, Serial Numbers: 87CA63061; UDI/DI 01008406821408432187CY6303211191000, Serial Numbers: 87CY63032; UDI/DI To be provided, Serial Numbers: 87CB63075, 87CA63058, 87CB63070, 87CB63069, 87CB63071, 87CB63077, 87CA63063, 87CB63064, 87CB63078, 87CA63062, 87CB63067, 87CA63057, 87CA63060, 87CB63065, 87CB63068, 87CB63074; UDI/DI Not Applicable, Serial Numbers: 87CY63019, 87CA63054, 87CZ63043, 87CA63055, 87CY63028, 87CY63037, 87CZ63044, 87CZ63045, 87CZ63048, 87CZ63041, 87CY63024, 87CY63017, 87CX63007, 87CX63008, 87CX63006, 87CX63014, 87CY63022, 87CX63015, 87CY63025, 87CX63010, 87CZ63052, 87CY63027, 87CY63026, 87CY63030, 87CZ63046, 87CX63013, 87CY63036, 87CZ63047, 87CY63033, 87CY63016, 87CX63005, 87CZ63039, 87CX63002, 87CX63012, 87CX63001, 87CY63023, 87CY63018; b) D670 CZT Gantry 7.25mm: Catalog number H3906DD, UDI/DI Not Applicable, Serial Numbers: CZGA55003, CZGA55004, CZGY55001; , UDI/DI To be provided, Serial Numbers: CZGB55006, CZGA55005, CZGB55007; c) DISCOV NM/CT 870 CZT: Catalog number H3906CW, UDI/DI 01008406821408432187CY6303811191200, Serial Numbers: 87CY63038; d) Discovery NMCT 670 CZT O540+: Catalog number H3906DD , UDI/DI Not Applicable, Serial Numbers: CZGZ55002; e) NM/CT 870 CZT CYBER DEFENCE: Catalog number H3906FK, UDI/DI To be provided, Serial Numbers: 87DB82001; f) NM/CT 870 CZT O540: Catalog number H3906CW , UDI/DI Not Applicable, Serial Numbers: 87CY63021, 87CZ63050, 87CZ63049, 87CZ63051, 87CY63034; g) NM/CT 870 CZT WSO: Catalog number 5376204-70-20, UDI/DI 01008406821408432187CA6305911210800, Serial Numbers: 87CA63059; UDI/DI 01008406821408432187CB6307611221000, Serial Numbers: 87CB63076; UDI/DI 01008406821408432187CB6307311220800, Serial Numbers: 87CB63073; : Catalog number H3906CW, UDI/DI To be provided, Serial Numbers: 87CB63066; : Catalog number To be provided, UDI/DI To be provided, Serial Numbers: 87CA63056; h) NM/CT 870 CZT: Catalog number H3906CW, UDI/DI Not Applicable, Serial Numbers: 87CY63031; i) NM/CT 870 CZT O540+ 7.25MM: Catalog number H3906CV, UDI/DI 01008406821408432187CX6300311180700, Serial Numbers: 87CX63003; UDI/DI 01008406821408432187CY6302911190800, Serial Numbers: 87CY63029

Distribution Pattern:

Worldwide distribution. US nationwide including Puerto Rico, Algeria, Argentina, Australia, Austria, Bahrain, Belgium, Bolivia, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Egypt, El Salvador, Ethiopia, Finland, France, Germany, Ghana, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Republic of Korea, Kuwait, Libya, Lithuania, Malaysia, Martinique, Mexico, Morocco, Netherlands, New Zealand, Nigeria, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom, Uzbekistan, and Vietnam.

Voluntary or Mandated:

Voluntary: Firm initiated