GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING: Medical Device Recall in 2025 - (Recall #: Z-0291-2026)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2025.

Data Source: FDA.
Product Description:

Elscint Model 4XX; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging

Product Classification:

Class II

Date Initiated: September 12, 2025
Date Posted: October 29, 2025
Recall Number: Z-0291-2026
Event ID: 97647
Reason for Recall:

GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.

Status: Ongoing
Product Quantity: 64 units
Code Information:

all serial numbers in distribution

Distribution Pattern:

Worldwide

Voluntary or Mandated:

Voluntary: Firm initiated