GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING: Medical Device Recall in 2025 - (Recall #: Z-2098-2025)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2025.
Data Source: FDA.
Product Description:
Varicam, Millennium VG, Millennium VG Hawkeye, Discovery VH Nuclear Medicine System
Product Classification:
Class II
Date Initiated: June 20, 2025
Date Posted: July 16, 2025
Recall Number: Z-2098-2025
Event ID: 97193
Reason for Recall:
Unintended radial detector motion may occur during patient setup or during patient scan if system does not have correct version of gantry software installed. Unintended detector motion may result in life-threatening injury.
Status: Ongoing
Product Quantity: 21
Code Information:
Serial Numbers: Millennium VG 1074; "Varicam Millennium VG Discovery VH" 1003 1092 10147 10242 10374 10378 10539 10543 10550 10553 10561 10569 10570 10571 10579 10582 10587 10599 10604 3101-1502
Distribution Pattern:
Worldwide - US Nationwide distribution.
Voluntary or Mandated:
Voluntary: Firm initiated