GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING: Medical Device Recall in 2026 - (Recall #: Z-1045-2026)
See the recall detail below. You can also see other recalls from the same firm in 2026.
SMV DSI, System, Tomography, Computed, Emission
Class II
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service (herein referred to as Impacted Systems and listed in the Affected Product Details section of this letter) could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury. There have been no reports of detector falls or injuries as a result of this potential issue. The Impacted Systems affected by this notice are those that have not been manufactured for several years. GE HealthCare has previously communicated End of Guaranteed Service for these Impacted Systems.
UDI/DI Not applicable: All serial numbers in distribution
US, Brazil, Egypt, Canada, Mexico, Turkey, Italy, Sweden, Switzerland, France, Spain Chile, Germany, Iran, Greece
Voluntary: Firm initiated