GE Medical Systems, LLC: Medical Device Recall in 2015 - (Recall #: Z-0196-2016)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2015.

Data Source: FDA.
Product Description:

MR Surgical Suite II Table, Model Number(s): M8074SS, M8053SS, M0074SS. UPDATE 11-4-2015: Model M0001SS

Product Classification:

Class II

Date Initiated: September 23, 2015
Date Posted: November 11, 2015
Recall Number: Z-0196-2016
Event ID: 72448
Reason for Recall:

GE became aware of an issue with the table transfer release mechanism of the MR Surgical Suite Option. The issue may prevent patient transfer between the MR and surgical tables.

Status: Terminated
Product Quantity: 23 (11 US; 12 OUS) Update 11-4-2015: 35 (14 US; 21 OUS)
Code Information:

Mfg Lot or Serial # System ID 00000304124MR1 904202MR8 00000000UA0116 330344MR750W 00000000UA0117 616267IMRI UA0326 614293MR9 00000307553MR8 415UCBHMR1 00000000UA0296 309655MR750W 00000000UA0396 713792AMR14 00000012228M63 608263MR4IOR N/A 614RMHTMR1 N/A 720848MR4 N/A 309655IMRI 00000299229MR5 AH5824MR01 00000014131M67 MRHM0323 00000018480M64 MR307144MR6 00000021158M61 MRHM0885 00000013254M68 082445160014 00001309FMM07L T4185506 DUMFMI60876024 M347028301 00000021695M62 M4194477 00000305145MR5 UC2547MR01 00001219FMM0H1 EM0105 00001105XF5047 PL2853MR01 00001251FMM044 34368MRS01 UPDATE 11-4-2015: Mfg Lot or Serial # System ID 2 602406IMR 3 507284NMRV N/A 214820ITABLE N/A 416369MR3T 000000PTST0008 082427040101 000000PTST0007 082427050024 000000PTST0003 CS1020MR01 N/A CS1006MR03 4 YM1733 000000PTST0010 YM3833 000000PTST0006 ZA2169MR01 000000PTST0005 10590MRS01

Distribution Pattern:

Worldwide Distribution - US Distribution including the states of : CA, CO, FL, IL, MI, OH, TX and WI., and to the countries of : AUSTRIA, BRAZIL, FRANCE, GERMANY, JAPAN, POLAND and TURKEY. UPDATE 11-4-2015: Additional states: AZ, MN. Additional countries: Canada, China, Czech Republic, South Africa, and United Kingdom

Voluntary or Mandated:

Voluntary: Firm initiated