GE Medical Systems, LLC: Medical Device Recall in 2025 - (Recall #: Z-2133-2025)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2025.
Data Source: FDA.
Product Description:
SIGNA PET/MR
Product Classification:
Class II
Date Initiated: May 23, 2025
Date Posted: July 30, 2025
Recall Number: Z-2133-2025
Event ID: 97030
Reason for Recall:
GE HealthCare has become aware that gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning. As a result, acoustic levels can exceed 99dB, the limit established by International Electrotechnical Commission (IEC 60601-2-33) when using hearing protection with a Noise Reduction Rating (NRR) of 29dB, as currently required in the Operator Manual.
Status: Ongoing
Product Quantity: N/A
Code Information:
GTIN: 00840682105378
Distribution Pattern:
Worldwide distribution - US Nationwide and the country of China.
Voluntary or Mandated:
Voluntary: Firm initiated