GE Medical Systems, SCS: Medical Device Recall in 2024 - (Recall #: Z-1781-2024)
See the recall detail below. You can also see other recalls from the same firm in 2024.
GE Healthcare Allia IGS 7 OR, Interventional fluoroscopic x-ray systems
Class II
GE HealthCare has become aware of the potential that adhesive material might not have been applied to some bolts of certain Innova IGS, Discovery IGS, Interact IGS, Allia IGS and Allia Pulse IGS products. This could result in a potential fall of the Detector/Elevator or X-ray tube of the frontal Gantry assembly.
UDI/DI 00195278215550, Serial Numbers: D3-23-026, D4-24-002
Worldwide distribution - US Nationwide in the states of CA, FL, GA, IN, MA, NC, OK, SC, TN, TX, VA and the countries of Armenia, China, France, Hungary, India, Italy, Japan, Kazakhstan, Lithuania, Mexico, Russia, Spain, Taiwan, Turkey, Vietnam
Voluntary: Firm initiated