GE OEC Medical Systems, Inc: Medical Device Recall in 2012 - (Recall #: Z-2136-2012)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2012.

Data Source: FDA.
Product Description:

Radiological Image Processing System The system is an aid to locate anatomical structures during open or percutaneous surgical procedures. It is indicated for use in medical conditions that may benefit from the use of stereotactic surgical technique. The system provides a reference to rigid anatomical structures such as sinus, skull, long bone, or vertebra, which are visible on medical images such as CS, MR, or X-ray.

Product Classification:

Class I

Date Initiated: April 2, 2007
Date Posted: August 15, 2012
Recall Number: Z-2136-2012
Event ID: 62414
Reason for Recall:

The FluoroTrak Spinal Navigation Application on the OEC 9900 EliteNAV could result in an incorrect position of the navigated instrument(s) versus the displayed reference image.

Status: Terminated
Product Quantity: 17
Code Information:

E9-0002-G, ES-0013-G, E9-0036-G, E9-0040-G, E9-0041-G, ES-0014-G, ES-0016-G, ES-0033-G, ES-0043-G, ES-0041-G, ES-0042-G, ES-OO84-G, ES-0086-G, ES-0163-G, E9-0089-G, ES-0026-G, ES-0044-G

Distribution Pattern:

Worldwide Distribution - USA including MO, IL, LA, FL, NH, TX, IN, PA, NJ and the countries of France and China

Voluntary or Mandated:

Voluntary: Firm initiated