Getinge Disinfection Ab: Medical Device Recall in 2025 - (Recall #: Z-0016-2026)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2025.

Data Source: FDA.
Product Description:

Getinge 88-Series Washer- Disinfector. Model Number: 88-5.

Product Classification:

Class II

Date Initiated: August 29, 2025
Date Posted: October 8, 2025
Recall Number: Z-0016-2026
Event ID: 97535
Reason for Recall:

Potential for device to overheat during operation if the circulation pressure calibration was not performed or had been performed incorrectly, and the low-pressure alarm is not triggered.

Status: Ongoing
Product Quantity: 902 units
Code Information:

Model Number: 88-5. UDI-DI Numbers: 07340153700109, 07340153700116. Software Version: GD 14080 Rev A. All Lot Numbers with electrical heating.

Distribution Pattern:

Worldwide - US Nationwide distribution in the states of Colorado, Massachusetts, New Mexico, New Hampshire and the countries of Algeria, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Czech Republic, Denmark, Estonia, Finland, France, Germany, Austria, Hong Kong, Hungary, India, Iraq, Ireland, Israel, Italy, Japan, Kenya, Korea, Republic of, Kuwait, Lithuania, Mexico, Netherlands, Norway, New Zealand, Poland, Qatar, Romania, Saudi Arabia, Singapore, Spain, Sweden, Thailand, Turkey, United Arab Emirates, United Kingdom.

Voluntary or Mandated:

Voluntary: Firm initiated