GETINGE US SALES LLC: Medical Device Recall in 2019 - (Recall #: Z-0987-2019)
See the recall detail below. You can also see other recalls from the same firm in 2019.
Maquet XS Flat Screen Monitor Holder, XS32 SPE, Configuration Part Number ARD22201040A Product Usage: This product is designed to support one or two medical grade flat screen(s) alone or in association with surgical lights.
Class II
The affected flat screen monitor holders might not be compliant with grounding requirements. As a consequence, there is a risk of mild to severe/fatal electric shock under certain conditions while using the flat screen monitor holder with voltage of 100-220V.
Serial Numbers 500009 500010 500011 500012 500013 500014
The products were distributed to the following US states: AZ and TX.
Voluntary: Firm initiated