Product Description:

Maquet XS Flat Screen Monitor Holder, XS32 SPE, Configuration Part Number ARD22201040A Product Usage: This product is designed to support one or two medical grade flat screen(s) alone or in association with surgical lights.

Product Classification:

Class II

Date Initiated: January 8, 2019
Date Posted: March 13, 2019
Recall Number: Z-0987-2019
Event ID: 82074
Reason for Recall:

The affected flat screen monitor holders might not be compliant with grounding requirements. As a consequence, there is a risk of mild to severe/fatal electric shock under certain conditions while using the flat screen monitor holder with voltage of 100-220V.

Status: Terminated
Product Quantity: 6
Code Information:

Serial Numbers 500009 500010 500011 500012 500013 500014

Distribution Pattern:

The products were distributed to the following US states: AZ and TX.

Voluntary or Mandated:

Voluntary: Firm initiated